A very compact, low-noise, fast-switching microwave signal generator series for OEM integration. Continuous coverage from 100 kHz (9 kHz with option 9K) up to 12, 20, or 26 GHz, with 0.001 Hz resolution, accurately leveled output power, and a full analog modulation suite.
Specifications reproduced from Model 845-OEM datasheet V2.59 (dc-81405202). Warranted at 23 ±5 °C after 30-minute warm-up unless stated.
1Introduction
Model 845-OEM board-level microwave signal generator module, shown with and without the RF shielding cans.
Very compact, portable analog signal generator models with a range from 9 kHz (with option 9K) up to 26.5 GHz. The Model 845-OEM is a series of low-noise and fast-switching microwave signal generators covering a continuous frequency range from as low as 100 kHz up to 12, 20, and 26 GHz, respectively, with a 0.001 Hz resolution.
The Model 845-OEM provides an accurately leveled output power range and high spurious suppression. Advanced frequency synthesis with fractional-N divider makes for low SSB phase noise and micro-Hz resolution.
The standard Model 845-OEM includes amplitude modulation (AM), DC-coupled low-distortion wideband frequency modulation (FM), PM, FSK and PSK, frequency chirp, and fast pulse modulation with internal pulse train generator. Three internal modulation sources are available, and all modulation modes can be combined. This allows the generation of complex modulation signals for modern communication and location systems. The combination of pulse modulation and FM simulates Doppler effects or chirp signals. Simultaneous AM and pulse modulation provide the types of signals occurring in pulse radar applications with rotating antenna. The combination of FM and AM can be used to check fading effects of FM receivers.
All Model 845-OEM units allow fast analog and digital sweeps including flexible list sweeps, where frequency, power, and dwell times can be set individually. A flexible triggering capability simplifies synchronization within test environments.
All units operate with an ultra-stable temperature-compensated 100 MHz reference (OCXO) to ensure minimal drift, and can be phase-locked to any stable external reference in a range from 1 to 250 MHz. Additionally, optimum phase-synchronous signals can be achieved by bypassing the internal reference and feeding a 100 MHz signal directly as reference. The 845-OEM supports various standard interfaces such as USB-TMC, LAN, and GPIB.
Definitions
The specifications describe the warranted performance of the instrument for 23 ±5 °C after a 30-minute warm-up period, unless otherwise stated.
Min/Max: Parameter range guaranteed by product design and/or production tested. Warranted performance specifications include guard-bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Typical: Expected mean values, not warranted performance.
Applications
R&D low-noise microwave source
Production testing (industry-leading switching times, high dynamic range)
Service and maintenance (battery operation)
Signal simulation (Radar, WiMax, UWB)
Aerospace & defense (pulse modulator, chirps)
Aerospace and defense: radar and EW signal simulation.SATCOM and radar receiver test.
2Available Options
Option
Description
HP
Delivers higher maximum output power, to a level up to +27 dBm.
PE2
Optional power level extension to accurately level below -120 dBm.
FS
Substantially enhances the switching speed.
B3
Adds an internal rechargeable battery module.
9K
Low-frequency extension down to 9 kHz (845-20, 845-12 only).
FILT
Harmonic filtering.
3Frequency Parameters & Range
Parameter
Min
Typical
Max
Note
Frequency range
100 kHz
12.0 GHz
845-OEM-12
100 kHz
20.0 GHz
845-OEM-20, settable to 20.5 GHz
100 kHz
26.0 GHz
845-OEM-26, settable to 30 GHz
9 kHz
Option 9 kHz
Resolution
0.001 Hz
Phase resolution
0.1 deg
Frequency / amplitude setting time
200 µs
300 µs
30 µs
Option FS
30 µs
Fswp < 2.2 GHz, Option NEC
4Level Performance
Figure 1: Typical maximum output power, 9 kHz to 1 MHz (option 9K). Figure 2: Maximum output power, 845-26, HP with and without option PE2. Typical, not guaranteed.
Output power. Check maximum output power plots. Level resolution is 0.01 dB (typical).
Configuration
Min
Max
Frequency range
Standard
-20 dBm
+15 dBm
9 kHz to 100 kHz (option 9K)
-20 dBm
+15 dBm
100 kHz to 26 GHz
Option PE
-90 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-90 dBm
+13 dBm
100 kHz to 26 GHz
Option PE2
-120 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-120 dBm
+13 dBm
100 kHz to 26 GHz
Option HP
-20 dBm
+15 dBm
9 kHz to 100 kHz (option 9K)
-20 dBm
+20 dBm
100 kHz to 10 MHz
-20 dBm
+21 dBm
0.01 to 12 GHz
-20 dBm
+20 dBm
12 to 20 GHz
-20 dBm
+19 dBm
20 to 24 GHz
-20 dBm
+15 dBm
24 to 26 GHz
Option HP and FILT
-20 dBm
+15 dBm
9 kHz to 100 kHz (option 9K)
-20 dBm
+19 dBm
100 kHz to 10 MHz
-20 dBm
+20 dBm
0.01 to 12 GHz
-20 dBm
+19 dBm
12 to 20 GHz
-20 dBm
+18 dBm
20 to 24 GHz
-20 dBm
+14 dBm
24 to 26 GHz
Option HP and PE2
-120 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-120 dBm
+20 dBm
100 kHz to 10 MHz
-120 dBm
+21 dBm
0.01 to 12 GHz
-120 dBm
+20 dBm
12 to 20 GHz
-120 dBm
+18 dBm
20 to 24 GHz
-120 dBm
+14 dBm
24 to 26 GHz
Options HP and PE2 + FILT
-120 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-120 dBm
+19 dBm
100 kHz to 10 MHz
-120 dBm
+20 dBm
0.01 to 12 GHz
-120 dBm
+18 dBm
12 to 20 GHz
-120 dBm
+17 dBm
20 to 24 GHz
-120 dBm
+13 dBm
24 to 26 GHz
Options HP and PE
-90 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-90 dBm
+20 dBm
100 kHz to 10 MHz
-90 dBm
+21 dBm
0.01 to 12 GHz
-90 dBm
+19 dBm
12 to 20 GHz
-90 dBm
+18 dBm
20 to 24 GHz
-90 dBm
+14 dBm
24 to 26 GHz
Option HP and PE + FILT
-90 dBm
+13 dBm
9 kHz to 100 kHz (option 9K)
-90 dBm
+19 dBm
100 kHz to 10 MHz
-90 dBm
+20 dBm
0.01 to 12 GHz
-90 dBm
+18 dBm
12 to 20 GHz
-90 dBm
+17 dBm
20 to 24 GHz
-90 dBm
+13 dBm
24 to 26 GHz
5Power Uncertainty & VSWR
Reverse Power Protection and VSWR
Figure 3: Typical VSWR (845-20). Typical, not guaranteed.
Parameter
Min
Typical
Max
Note
Reverse power protection — DC voltage
±15 V
Reverse power protection — RF power
+30 dBm
Output impedance
50 Ω
VSWR
1.5
< 20 GHz
2.0
> 20 GHz
Power Level Uncertainty (ALC on, Model 845-OEM + PE2/PE)
Figures 4 and 5: Typical frequency response, 0 to 20 GHz at -20, 0, and +20 dBm. Typical, not guaranteed.
Values are the warranted figure, with the typical value in parentheses.
Frequency
>15 dBm to Pmax
-15 to +15 dBm
-70 to -15 dBm
< -70 dBm
10 MHz to 10 GHz
1.6 dB (0.8 dB)
1.0 dB (0.3 dB)
1.5 dB (0.4 dB)
1.8 dB (0.5 dB)
100 Hz to fmax
1.6 dB (0.8 dB)
1.0 dB (0.3 dB)
1.5 dB (0.4 dB)
2.0 dB (0.5 dB)
User flatness correction: up to 2000 points. Temperature effects: 0.015 dB/°C (typical) in the range 0 to 45 °C.
6Phase Noise
Figure 6: Phase noise performance (1 Hz to 100 MHz offset) at different carrier frequencies. Typical, not guaranteed.
SSB phase noise (standard)
Typical
Max
1 GHz, 10 Hz offset
-80 dBc/Hz
-77 dBc/Hz
1 GHz, 20 kHz offset
-130 dBc/Hz
-128 dBc/Hz
1 GHz, 100 kHz offset
-131 dBc/Hz
-129 dBc/Hz
4 GHz, 10 Hz offset
-68 dBc/Hz
-65 dBc/Hz
4 GHz, 20 kHz offset
-118 dBc/Hz
-116 dBc/Hz
4 GHz, 100 kHz offset
-119 dBc/Hz
-117 dBc/Hz
20 GHz, 10 Hz offset
-51 dBc/Hz
-48 dBc/Hz
20 GHz, 20 kHz offset
-104 dBc/Hz
-102 dBc/Hz
20 GHz, 100 kHz offset
-104 dBc/Hz
-100 dBc/Hz
Wideband noise
-150 dBc/Hz
Amplitude noise at 10 GHz
-130 dBc/Hz (-140 dBm)
7Spectral Purity
Figure 7: Harmonics, standard. Figure 8: Harmonics, 845-12 with option FILT. Figure 9: Harmonics, 845-20 with option FILT. Figure 10: Harmonics, 845-26 with option FILT. Typical, not guaranteed.
Spectral purity at +5 dBm.
Parameter
Typical
Max
Note
Output harmonics
-40 dBc
-30 dBc
See plot
845-OEM-12 with option FILT, 10 MHz to 8 GHz
-40 dBc
-30 dBc
See plot
845-OEM-12 with option FILT, 8 to 12 GHz
-60 dBc
-50 dBc
845-OEM-20 with option FILT, 10 MHz to 12 GHz
-40 dBc
-30 dBc
See plot
845-OEM-20 with option FILT, 12 to 20 GHz
-65 dBc
-50 dBc
845-OEM-26 with option FILT, 10 MHz to 15 GHz
-40 dBc
-30 dBc
See plot
845-OEM-26 with option FILT, 15 to 20 GHz
-60 dBc
-50 dBc
Sub-harmonics
-75 dBc
-65 dBc
< 20 GHz
-50 dBc
-40 dBc
> 20 GHz
Non-harmonic spurious, < 312 MHz
-80 dBc
-66 dBc
CW +10 dBm, > 3 kHz offset
Non-harmonic spurious, > 312 to 625 MHz
-75 dBc
-70 dBc
Non-harmonic spurious, > 625 MHz to 1.5 GHz
-75 dBc
-65 dBc
Non-harmonic spurious, 1.5 to 2.5 GHz
-70 dBc
-65 dBc
Non-harmonic spurious, > 2.5 to 5 GHz
-65 dBc
-60 dBc
Non-harmonic spurious, > 5 to 10 GHz
-60 dBc
-55 dBc
Non-harmonic spurious, > 10 to 20 GHz
-55 dBc
-50 dBc
Non-harmonic spurious, > 20 GHz
-50 dBc
-45 dBc
Residual FM @ 10 GHz
15 Hz
0.3 kHz to 3 kHz, weighted (ITU-T), RMS
Residual AM @ 10 GHz
0.02%
RMS value (0.01 kHz to 15 kHz)
8Sweeping & Switching
Figure 15: Typical switching transient, 12 GHz to 14 GHz step. Typical, not guaranteed.
Digital power / frequency / list sweeps. Sweep type: linear, logarithmic, random. Sweeps can be performed with combined internal or external AM/FM/PM/pulse modulation running; with modulation enabled, the minimum step time increases to 2 ms.
Parameter
Min
Typical
Max
Note
Step time (tstep)
400 µs
19998 s
40 µs
Option FS
Dwell time (tdwell)
10 µs
9999 s
Off-time incl. transient time (toff)
0
9999 s
Transient time (ttrs)
270 µs
30 µs
Option FS
Timing delay (tdly)
2 to 10 µs
50 ns
Option FS
Time resolution
0.1 µs
5 ns
Option FS
Timing accuracy per point
3 µs
5 ns
Option FS
9Reference Frequency
REF IN input and REF OUT output are at the rear panel.
Parameter
Min
Typical
Max
Note
Internal reference frequency
100 MHz
Initial accuracy
±40 ppb
Calibrated at 23 ±3 °C at time of calibration, user adjustable
Temperature stability (0 to 50 °C)
±100 ppb
Aging, 1st year
0.5 ppm
Aging per day (after 30 days operation)
5 ppb
Warm-up time
5 min
Output of internal reference
10 / 100 MHz
Output power
0 dBm
Output impedance
50 Ω
Bypass internal reference input
100 MHz, -5 to +10 dBm
High phase-synchronous mode
External input range (phase lock)
1 MHz
250 MHz
User programmable
Reference input level
-5 dBm
0 dBm
+13 dBm
Lock range
±1.5 ppm
Reference input impedance
50 Ω
10Outputs & Triggers
Multi-Purpose Output (FUNC OUT)
Output is FUNC OUT at the rear panel.
Multifunction generator
Min
Typical
Max
Note
Frequency range (sine, triangle, square)
10 Hz
3 MHz
Sine
10 Hz
1 MHz
Triangle
50 kHz
Square
Frequency resolution
0.1 Hz
Output voltage amplitude (peak-peak)
10 mV
2 V
Sine, triangle
5 V
Square (CMOS output)
Harmonic distortion
1%
< 100 kHz, 1 Vpp
Output impedance
50 Ω / CMOS
Sine, triangle / square wave
Video output (of internal pulse modulator)
Min
Typical
Max
Output
CMOS
Period
30 ns
50 s
Pulse width
15 ns
50 s
RF delay
10 ns
TRIG OUT synchronization modes for multiple sources: trigger on sweep start, trigger on each point, signal valid (option FS).
Trigger (TRIG IN)
Trigger input / output timing: TRIG IN, TRIG OUT (signal valid), and RF valid.
Input is TRIG IN at the rear panel. Trigger Output (TRIG OUT) follows the Multi-Purpose Output (FUNC OUT) above.
Parameter
Min
Typical
Max
Note
Trigger types
Continuous, single, gated, gated direction
Trigger source
RF key, external, bus (GPIB, LAN, USB)
Trigger modes
Continuous free run, trigger and run, reset and run
Trigger latency
2 µs
130 ns
Option FS
Trigger uncertainty
5 µs
10 ns
Option FS
External trigger delay
50 µs
40 s
Programmable
50 ns
10 s
Option FS
External delay resolution
15 ns
10 ns
Option FS
Trigger modulo
1
255
Execute only on Nth trigger event
Trigger polarity
Rising, falling
External trigger input threshold
0.85 V
0.9 V
0.95 V
TTL compatible
External trigger input voltage range
-0.5 V
+5.5 V
TTL compatible
External trigger input hysteresis
60 mV
11Modulation
Combinations of AM/PM/FM/pulse are possible. See the user manual for more details.
Pulse Modulation
Parameter
Min
Typical
Max
Note
On/off ratio
75 dB (typical)
At +10 dBm
Repetition frequency
DC
10 MHz
Pulse width
100 ns
5 s
ALC hold
30 ns
5 s
ALC hold, options NP or FS
500 ns
ALC on
Pulse rise/fall time
30 ns
< 5 GHz
5 ns
> 5 GHz
Duty cycle
0.05%
99.95%
Pulse resolution
30 ns
Polarity
Selectable
External input threshold
0.85 V
0.9 V
0.95 V
TTL compatible
External input voltage range
-0.5 V
+5.5 V
TTL compatible
External input hysteresis
60 mV
Onset delay (to RF)
20 ns
40 ns
Pulse Pattern Modulation
Parameter
Min
Typical
Max
Note
Pulse bit width
100 ns
ALC hold
30 ns
ALC hold, option NP or FS
500 ns
ALC on. Uses internal pattern generator.
Programmable pattern length
2
4096
Duty cycle
0.05%
99.95%
Pulse bit resolution
30 ns
5 ns
Option NP or FS
Polarity
Selectable
N-Pulse Modulation (Option FS)
N-Pulse modulation timing: up to four pulses (P1–P4) with individual widths, delays, and off-times within the period T, referenced to the trigger and delay.
Parameter
Min
Typical
Max
Note
Period
30 ns
21 s
N-Pulse type
1 to 4
No overlapping with other pulses
N-Pulse width (P1, P2, P3, P4 widths)
30 ns
21 s
ALC hold
500 ns
21 s
ALC on
N-Pulse delay (P1 delay)
0 ns
21 s
N-Pulse delay (P2, P3, P4)
Min: end of prior pulse + min off time
N-Pulse width and delay resolution
5 ns
N-Pulse off time
1 µs
21 s
Determined by pulse delay and prior pulse width
Frequency Modulation
Parameter
Min
Typical
Max
Note
Maximum frequency deviation (peak)
0
0.05·f
< 0.5 GHz
Maximum frequency deviation (peak)
0
N·200 MHz
0.5 to 2.5 GHz (N=0.125); 2.5 to 5 GHz (N=0.25); 5 to 10 GHz (N=0.5); > 10 to 20 GHz (N=1)
Deviation accuracy, < 100 kHz rate
0.5%
2%
Deviation accuracy, > 100 kHz rate
2%
5%
Distortion
< 1%
1 kHz rate, 50 kHz deviation
Modulation waveforms / rate
DC
Sine, triangle, FSK
100 kHz
> -3 dB frequency response
External input sensitivity (AC coupled)
0 to N·200 MHz / V
Adjustable for ±1 V range
External input sensitivity (DC coupled)
0 to N·100 MHz / V
Discrete values, ±5 V range
Total harmonic distortion
< 1%
1 kHz rate & N·1 MHz deviation
Frequency Chirps (linear ramp, up/down)
Parameter
Min
Max
Note
Span
10%
Of carrier frequency
Chirp time (tchirp)
10 ns
60 s
Total duration of finite repeated chirps (tchirp · repetitions)
64.1 s
21.4 s
With option FS
Slope
100 MHz/µs
Number of frequencies
65,000
Phase Modulation
Parameter
Min
Typical
Max
Note
Phase deviation (peak)
0
N·300 rad
Modulation rate
DC
100 kHz
> -3 dB frequency response; max. phase deviation degrades above 20 kHz modulation rate
Modulation waveforms
Sine, triangle, FSK
External input sensitivity
Settable 0.1 rad/V to 360 rad/V
Total harmonic distortion
< 1%
1 kHz rate & N·100 rad deviation
Amplitude Modulation
Parameter
Min
Typical
Max
Note
Modulation rate
0.1 Hz
50 kHz
Modulation waveforms
Sine, triangle, square
Modulation depth
0%
100%
Settable
Accuracy (f ≥ 10 MHz)
0.8%
1.4%
f-carrier, modulation depth < 80%
Distortion (f ≥ 10 MHz)
1%
2%
& 1 kHz modulation rate, power 0 dBm
External input range
0 V
5 V
To GND
12Ordering Information
Host model
Product / option
Description
845-OEM-12
845-OEM-12
12 GHz microwave signal generator
845-OEM-20
845-OEM-20
20 GHz microwave signal generator
845-OEM-26
845-OEM-26
26 GHz microwave signal generator
845-OEM-12/20
Option 9K
Frequency range extension to 9 kHz
845-OEM-26
Option PE
Mechanical step attenuator (26 GHz version) down to -90 dBm
845-OEM-XX
Option PE2
Mechanical step attenuator to -120 dBm
845-OEM-XX
Option HP
Higher output power
845-OEM-12/20/26
Option FILT
Improved harmonic filtering
845-OEM-XX
Option FS
Ultra-fast switching speed (export control)
845-OEM-XX
Option Retrofit
Applies when options are added after first delivery
845-OEM-XX
Option NP
Narrow pulse (30 ns minimum PW and 5 ns resolution)
For a configured quote, contact Berkeley Nucleonics Corporation · info@berkeleynucleonics.com · 415-453-9955.
13General Characteristics
Characteristic
Specification
Remote programming interfaces
Ethernet 100BaseT LAN interface; USB 2.0 host & device; GPIB (IEEE-488.2, 1987) with listen and talk (optional)
Control language
SCPI version 1999.0
Power requirements
6.25 ±0.2 VDC; 20 W maximum
Main adapter supplied
100–240 VAC in / 6 V, 6.0 A DC out
Environmental
Levels similar to MIL-PRF-28800F Class 3 & 4. Samples type tested to be robust against the environmental stresses of storage, transportation, and end-use (temperature, humidity, shock, vibration, altitude, and power line conditions).
Operating temperature range
0 to 45 °C
Storage temperature range
-40 to 70 °C
Max. relative humidity
85% up to 45 °C ambient
Operating and storage altitude
Up to 15,000 feet (4600 m)
EMC
Complies with EMC regulations and directives for emission and immunity to interference (EN 61326-1 Industrial, EN/IEC 61326-2-1)
Safety
Complies with applicable safety regulation in line with IEC/EN 61010-1
Weight
≤ 6 lbs (2.5 kg) net, ≤ 8 lb (4 kg) shipping
Dimensions (H x W x L)
4.21 in H x 6.77 in W x 10.63 in L incl. connectors [106 mm H x 172 mm W x 270 mm L]
Recommended calibration cycle
24 months
EMC complies with EMC regulations and directives for emission and immunity to interference (EN 61326-1 Industrial, EN/IEC 61326-2-1). Safety complies with applicable safety regulation in line with IEC/EN 61010-1.